AF849 - Autoclave sk07-b with printer

Autoclave SK07-B carries out only type B cycles (repeated vacuum stages for...

Enter to visualize prices and buy

Autoclave SK07-B carries out only type B cycles (repeated vacuum stages for the sterilisation of solid objects, hollow bodies, cannules and porous material both pouched or not) and complies with the EN 13060 standard. Its last generation electronics and high quality components grant the best possible performance with waste of energy and without impairing the sterilised materials.
• Limited water consumption: only 220 ml for the 134°C standard cycle
• Limited electric consumption: less than 1100W for the 134°C standard cycle

Conceived to optimise operator's time: • Easy to use: LCD display with menu images and icons.
• History archive of the cycles performed.
• Customised sterilising cycles repetition of the mostly used cycle.
• Signal icons.
• Space-saving: the autoclave can be integrated into an autoclave unit.
• Automatic Last Advanced Cycle.
• Automatic Stand-By at low temperature.
• Completely digital.
• Interface with PC for diagnostics and maintenance.

Functional, modern and elegant design:
• Highly integrated SMT electronics.
• Clock and calendar functions integrated.
• The autoclave can be connected to the main water supply thanks to its demineralization system.
• Double head vacuum pump.
• High performance vacuum pump
• Platinum thermal sounds, precise and reliable.V • USB port for connection to a PC.
• Good accessibility for component servicing.
• Adjustable front / lateral feet, distance between them: 36 cm x 29 cm
• two practical handles on the top (height of autoclave with handles cm 42.5)

Performances:
• Max load (solid pouched objects): 2 kg.
• Max load (non-pouched solid objects): 4 kg.
• Max porous load: 1,3 kg
• Autonomy of 10-12 cycles with inner tanks.
• Automatic tank load.

Performed cycles:
Standard 134°C Cycle
• 3-step fractioned vacuum of 13kPa (–870mbar).
• Sterilisation time 4 mins.
• Unload adn drying time 14 mins.
• Average time required for a complete cycle 37 mins. (depending on the loaded materials).
Standard 121°C cycle
• 3-step fractioned vacuum of 13kPa (–870mbar).
• Sterilisation time 16 mins.
• Unload and drying time 14 mins.
• Average time required for a complete cycle: 47 mins. (depending on the loaded materials).
Prion 134°C cycle
• 3-step fractioned vacuum of 13kPa (–870mbar).
• Sterilisation time 18 mins.
• Unload and drying time 14 mins
• Average time required for a complete cycle 53 mins. (depending on the loaded materials).

Test programms:
• Vacuum Test.
• Bowie & Dick Test
• 134°C Helix Test
• 121°C Helix Test

Technical features:
• Voltage: 230 V
• Power: 1725 W
• Max. absorbed current 7,5 A
• Sterilisation chamber's capacity: 15 lt.
• Tank's capacity: 3.65 lt.
• Dimensions: Lenght 44,9 x Height 39 x Width 66 cm

Endowment
• 3 trays
• 1 pair of forceps to extract the tray
• 1 pipe to load/unload the tank
• 1 funnel
• External printer


Vacuum Tests:

Bowie&Dick Test Vapour penetration in porous material
134°C Helix Test Vapour penetration in cavities
121°C Helix Test Vapour penetration in cavities

background

Disclaimer

Disclaimer

In compliance with the new guidelines of 28/03/2013 from the Italian Health Department concerning advertising of Medical Devices, In vitro diagnostic medical devices, and surgical devices, we would like to inform users that the information contained in this website is exclusively meant to professional operators. Clicking on “I Accept” you declare to be a professional operator.