Vapour penetration test for Class B autoclaves – Bowie & Dick test, 1 piece

  • Bowie & Dick steam penetration test for Class B autoclaves
  • Enables verification of the vacuum conditions within the autoclave chamber
  • Identifies instances of autoclave malfunction
  • Pack of 1
Ref: AF860B


Log in to visualise prices and purchase

Prices on Tecniwork.it are visible after logging into the website reserved to professionals only.

Purchase securely

Thousands of satified customers

Quick delivery

in 2/3 working days
Product description

Disposable Bowie & Dick test strips, supplied in a ready-to-use pack, for assessing steam penetration in Class B autoclaves. 

Allows verification of air removal in a saturated steam autoclave with fractional vacuum and the corresponding vacuum conditions within the autoclave chamber.

It highlights cases of autoclave malfunction.

Following the vacuum stages, the chemical indicator in the Test Card changes colour to dark, indicating that air has been sufficiently removed to allow complete steam penetration. If residual air remains trapped during the steam exposure phase, the colour change will be incomplete and non-uniform.


Frequency:

- Daily

- After any repair or maintenance work

- When the autoclave is moved.


Complies with ISO 11140-1 and ISO 11140-4 for the Type 2 section. 

The indicator located on the outside of the package complies with ISO 11140-1 Type 1.


Pack of 1

Details

Operating conditions: 134°C for 3.5 minutes or 121°C for 15 minutes 

Colour change indication: from pink to dark.

Hints

Instructions for use:

- Place the B&D Test Pack inside the empty sterilisation chamber on the lower shelf below the drain, with the chemical indicator strips on the sheet facing upwards.

- Perform the test with the chamber empty; the test is accurate when performed with an empty chamber. Do not use the cycle to sterilise any devices or packs.

- Start the sterilisation or test cycle if the autoclave has a specific one. Remove the B&D Test Pack from the autoclave.

- Open the B&D Test Pack and examine the B&D Test Sheet. The test is considered positive if the chemical indicator has changed colour uniformly and consistently to the prescribed shade. The autoclave cycle phase is functioning correctly.

- Record the information and retain the test.